Manual for BP200 - Troubleshooting and error codes + PDF Download

Manual for BP200 - Troubleshooting and error codes + PDF Download

Brand: ReliOn

Category: Blood Pressure Monitor

Manual Type:

Language:

Uploaded: April 8, 2025, 1:43 p.m.

Warranty Summary

ReliOn warrants the BP200 monitor for 5 years and the arm cuff for 1 year from purchase date against defects in materials and workmanship.
  1. Sit and rest for 5 minutes before measurement.
  2. Apply the arm cuff correctly on your left or right arm, 1-2 cm above the elbow.
  3. Select User ID (A or B).
  4. Press Start/Stop. The cuff will inflate automatically.
  5. Remain still during measurement.
  6. Reading will be displayed after deflation.
  7. Remove cuff and turn off the monitor.

Troubleshooting and error codes

SYMBOL CAUSE CORRECTION
Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in "TAKING A MEASUREMENT". If this error continues to appear, contact your physician.
Movement during measurement. Remain still and do not talk during the measurement. Carefully read and repeat the steps in "TAKING A MEASUREMENT".
Batteries are low. Recommend to replace the batteries with new ones ahead of time. Refer to "BATTERY INSTALLATION".
Batteries are depleted. Replace the four batteries. Refer to "BATTERY INSTALLATION".
E1 Air plug disconnected. Insert the plug securely. Refer to "APPLYING THE ARM CUFF".
E1 Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to "APPLYING THE ARM CUFF".
E1 Air is leaking from the arm cuff. Replace the cuff with a new one.
E2 Cuff over or under inflated. Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT". If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to "TAKING A MEASUREMENT".
E3 The arm cuff was inflated above 299 mmHg when inflating the cuff manually. Do not inflate the cuff above 299 mmHg. Refer to "TAKING A MEASUREMENT".
E4 Movement during measurement. Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT".
E5 Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to "APPLYING THE ARM CUFF".
Er Device error. Contact Customer Service.

Troubleshooting and error codes

SYMBOL CAUSE CORRECTION
Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in "TAKING A MEASUREMENT". If this error continues to appear, contact your physician.
Movement during measurement. Remain still and do not talk during the measurement. Carefully read and repeat the steps in "TAKING A MEASUREMENT".
Batteries are low. Recommend to replace the batteries with new ones ahead of time. Refer to "BATTERY INSTALLATION".
Batteries are depleted. Replace the four batteries. Refer to "BATTERY INSTALLATION".
E1 Air plug disconnected. Insert the plug securely. Refer to "APPLYING THE ARM CUFF".
E1 Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to "APPLYING THE ARM CUFF".
E1 Air is leaking from the arm cuff. Replace the cuff with a new one.
E2 Cuff over or under inflated. Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT". If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to "TAKING A MEASUREMENT".
E3 The arm cuff was inflated above 299 mmHg when inflating the cuff manually. Do not inflate the cuff above 299 mmHg. Refer to "TAKING A MEASUREMENT".
E4 Movement during measurement. Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT".
E5 Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to "APPLYING THE ARM CUFF".
Er Device error. Contact Customer Service.