Warranty Summary
SYMBOL | CAUSE | CORRECTION |
---|---|---|
Irregular heartbeats are detected. | Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in "TAKING A MEASUREMENT". If this error continues to appear, contact your physician. | |
Movement during measurement. | Remain still and do not talk during the measurement. Carefully read and repeat the steps in "TAKING A MEASUREMENT". | |
Batteries are low. | Recommend to replace the batteries with new ones ahead of time. Refer to "BATTERY INSTALLATION". | |
Batteries are depleted. | Replace the four batteries. Refer to "BATTERY INSTALLATION". | |
E1 | Air plug disconnected. | Insert the plug securely. Refer to "APPLYING THE ARM CUFF". |
E1 | Arm cuff not applied correctly. | Apply the arm cuff correctly. Refer to "APPLYING THE ARM CUFF". |
E1 | Air is leaking from the arm cuff. | Replace the cuff with a new one. |
E2 | Cuff over or under inflated. | Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT". If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to "TAKING A MEASUREMENT". |
E3 | The arm cuff was inflated above 299 mmHg when inflating the cuff manually. | Do not inflate the cuff above 299 mmHg. Refer to "TAKING A MEASUREMENT". |
E4 | Movement during measurement. | Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT". |
E5 | Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. Refer to "APPLYING THE ARM CUFF". |
Er | Device error. | Contact Customer Service. |
SYMBOL | CAUSE | CORRECTION |
---|---|---|
Irregular heartbeats are detected. | Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in "TAKING A MEASUREMENT". If this error continues to appear, contact your physician. | |
Movement during measurement. | Remain still and do not talk during the measurement. Carefully read and repeat the steps in "TAKING A MEASUREMENT". | |
Batteries are low. | Recommend to replace the batteries with new ones ahead of time. Refer to "BATTERY INSTALLATION". | |
Batteries are depleted. | Replace the four batteries. Refer to "BATTERY INSTALLATION". | |
E1 | Air plug disconnected. | Insert the plug securely. Refer to "APPLYING THE ARM CUFF". |
E1 | Arm cuff not applied correctly. | Apply the arm cuff correctly. Refer to "APPLYING THE ARM CUFF". |
E1 | Air is leaking from the arm cuff. | Replace the cuff with a new one. |
E2 | Cuff over or under inflated. | Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT". If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to "TAKING A MEASUREMENT". |
E3 | The arm cuff was inflated above 299 mmHg when inflating the cuff manually. | Do not inflate the cuff above 299 mmHg. Refer to "TAKING A MEASUREMENT". |
E4 | Movement during measurement. | Repeat measurement. Remain still and do not talk during measurement. Refer to "TAKING A MEASUREMENT". |
E5 | Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. Refer to "APPLYING THE ARM CUFF". |
Er | Device error. | Contact Customer Service. |