Warranty Summary

The CVS Health Advanced Wrist Blood Pressure Monitor is warrantied for the estimated life of the monitor against manufacturer defects.
  1. Prepare for Measurement: Sit relaxed for 5 minutes before measurement. Avoid eating, smoking and exertion.
  2. Apply Cuff: Place cuff on wrist, 1/2 inch from hand, with monitor face inside wrist. Secure snugly.
  3. Start Measurement: Press POWER. Cuff inflates, measurement starts automatically.
  4. View Results: Systolic, diastolic pressure and pulse are displayed after measurement.
  5. Stop Measurement: Press POWER button anytime to stop.

Troubleshooting and error codes

Error No. Possible cause(s) Remedy
ERR 1 No pulse was detected.* Ensure the cuff is properly placed on the inside of your wrist and repeat a measurement.
ERR 2 Unnatural pressure impulses influenced the measurement result. Reason: The arm was moved during the measurement (artefact). Repeat, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated. Re-position cuff and repeat the measurement.
ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures. Take another reading following directions carefully. Contact your doctor if you continue to get unusual readings.
HI The cuff pressure is too high. Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40). Repeat the measurement.*

Troubleshooting and error codes

Error No. Possible cause(s) Remedy
ERR 1 No pulse was detected.* Ensure the cuff is properly placed on the inside of your wrist and repeat a measurement.
ERR 2 Unnatural pressure impulses influenced the measurement result. Reason: The arm was moved during the measurement (artefact). Repeat, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated. Re-position cuff and repeat the measurement.
ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures. Take another reading following directions carefully. Contact your doctor if you continue to get unusual readings.
HI The cuff pressure is too high. Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40). Repeat the measurement.*