Manual for 400W SERIES - Troubleshooting and error codes + PDF Download

Manual for 400W SERIES - Troubleshooting and error codes + PDF Download

Brand: CVS Health

Category: blood pressure monitor

Manual Type:

Language:

Uploaded: April 8, 2025, 1:52 p.m.

Manual Publish Date: 2019-09-01

Warranty Summary

Limited lifetime warranty against manufacturer defects for the monitor and cuff.
  1. Prepare for measurement: avoid eating, smoking, exertion, relax for 5 minutes.
  2. Apply cuff to wrist, 1cm from hand, snug but not tight.
  3. Sit with arm supported at heart level.
  4. Press POWER button to start measurement.
  5. Remain still until measurement is complete.
  6. Read systolic, diastolic pressure and pulse on display.

Troubleshooting and error codes

Error No. Possible cause(s) / Remedy
ERR 1 No pulse was detected.* Ensure the cuff is properly placed on the inside of your wrist and repeat a measurement.
ERR 2 Unnatural pressure impulses influenced the measurement result. Reason: The arm was moved during the measurement (artefact). Repeat, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated. Re-position cuff and repeat the measurement.
ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures. Take another reading following directions carefully. Contact your doctor if you continue to get unusual readings.
HI The cuff pressure is too high. Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40). Repeat the measurement.*

Troubleshooting and error codes

Error No. Possible cause(s) / Remedy
ERR 1 No pulse was detected.* Ensure the cuff is properly placed on the inside of your wrist and repeat a measurement.
ERR 2 Unnatural pressure impulses influenced the measurement result. Reason: The arm was moved during the measurement (artefact). Repeat, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated. Re-position cuff and repeat the measurement.
ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures. Take another reading following directions carefully. Contact your doctor if you continue to get unusual readings.
HI The cuff pressure is too high. Relax for 5 minutes and repeat the measurement.*
LO The pulse is too low (less than 40). Repeat the measurement.*